CVE-2020-12012
https://notcve.org/view.php?id=CVE-2020-12012
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI. Baxter ExactaMix EM 2400 y EM 1200, Versiones ExactaMix EM2400 Versiones 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versiones 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versiones 1.10, 1.11 y 1.13, y ExactaMix EM1200 Versiones 1.1, 1.2 y 1.4, poseen credenciales embebidas de cuenta administrativa para la aplicación ExactaMix. Una explotación con éxito de esta vulnerabilidad puede permitir que un atacante con acceso físico consiga acceso no autorizado para visualizar y actualizar la configuración o los datos del sistema. • https://www.us-cert.gov/ics/advisories/icsma-20-170-01 • CWE-259: Use of Hard-coded Password CWE-798: Use of Hard-coded Credentials •
CVE-2020-12016
https://notcve.org/view.php?id=CVE-2020-12016
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI. Baxter ExactaMix EM 2400 y EM 1200, Versiones ExactaMix EM2400 Versiones 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versiones 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versiones 1.10, 1.11, 1.13, 1.14 y ExactaMix EM1200 Versiones 1.1, 1.2 , 1.4 y 1.5, poseen credenciales embebidas de cuenta administrativa para el sistema operativo de ExactaMix. Una explotación con éxito de esta vulnerabilidad puede permitir a un atacante que haya obtenido acceso no autorizado a los recursos del sistema, incluido el acceso para ejecutar software o para visualizar y actualizar archivos, directorios o la configuración del sistema. • https://www.us-cert.gov/ics/advisories/icsma-20-170-01 • CWE-259: Use of Hard-coded Password CWE-798: Use of Hard-coded Credentials •
CVE-2020-12020
https://notcve.org/view.php?id=CVE-2020-12020
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user. Baxter ExactaMix EM 2400 Versiones 1.10, 1.11 y 1.13 y ExactaMix EM1200 Versiones 1.1, 1.2 y 1.4, no restringe que los usuarios no administrativos consigan acceso al sistema operativo y editen el script de inicio de la aplicación. Una explotación con éxito de esta vulnerabilidad puede permitir a un atacante alterar el script de inicio como el usuario de acceso limitado • https://www.us-cert.gov/ics/advisories/icsma-20-170-01 • CWE-668: Exposure of Resource to Wrong Sphere •
CVE-2020-12024
https://notcve.org/view.php?id=CVE-2020-12024
Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI. Baxter ExactaMix EM 2400 versiones 1.10, 1.11, 1.13, 1.14 y ExactaMix EM1200 Versiones 1.1, 1.2, 1.4 y 1.5, no restringen el acceso a la interfaz USB de un usuario no autorizado con acceso físico. Una explotación con éxito de esta vulnerabilidad puede permitir a un atacante con acceso físico al sistema la capacidad de cargar una carga útil no autorizada o acceso no autorizado al disco duro mediante el arranque de un Sistema Operativo USB en vivo. • https://www.us-cert.gov/ics/advisories/icsma-20-170-01 • CWE-284: Improper Access Control •