5 results (0.014 seconds)

CVSS: 8.8EPSS: 0%CPEs: 3EXPL: 1

25 Aug 2021 — An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device. Una vulnerabilidad de saneamiento inapropiado de la entrada en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite que un atacante remoto no autenticado obtenga acceso a la línea de comandos a n... • https://www.bbraunusa.com/en.htm • CWE-134: Use of Externally-Controlled Format String •

CVSS: 10.0EPSS: 0%CPEs: 3EXPL: 1

25 Aug 2021 — An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets. Una vulnerabilidad de Verificación Insuficiente de la Autenticidad de Datos en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite a un atacante ... • https://www.bbraunusa.com/en.htm • CWE-347: Improper Verification of Cryptographic Signature •

CVSS: 9.1EPSS: 0%CPEs: 3EXPL: 1

25 Aug 2021 — An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten. Una vulnerabilidad de Carga no Restringida de Archivos de Tipo Peligroso en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite a atacantes remotos cargar cualquier archivo en el directorio /tmp del dispositivo mediante la API de la pági... • https://www.bbraunusa.com/en.htm • CWE-434: Unrestricted Upload of File with Dangerous Type •

CVSS: 7.5EPSS: 0%CPEs: 3EXPL: 1

25 Aug 2021 — A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration. Una vulnerabilidad de Transmisión en Texto sin Cifrar de Información Confidencial en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite a un atacante remoto conseguir información confidencial al espiar el tráfico de red. • https://www.bbraunusa.com/en.htm • CWE-319: Cleartext Transmission of Sensitive Information •

CVSS: 8.6EPSS: 0%CPEs: 3EXPL: 0

25 Aug 2021 — A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands. Una vulnerabilidad de Falta de Autenticación para una Función Crítica en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite a un atacante remoto reconfigurar el dispositivo desde una fuente desconocida debido a una falta de autenticación en los comandos ... • https://www.bbraunusa.com/en.htm • CWE-306: Missing Authentication for Critical Function •