Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Baxter SIGMA Spectrum Infusion System 6.05 (modelo 35700BAX) con un módulo de batería inalámbrica (WBM), en su versión 16, contiene una contraseña embebida, lo que proporciona acceso a información biomédica básica, opciones del dispositivo limitadas y la configuración de red del WBM, si aplica. La contraseña embebida podría permitir que un atacante con acceso físico al dispositivo acceda a las funciones de gestión para hacer cambios no autorizados en la configuración a las opciones biomédicas, como encender y apagar las conexiones inalámbricas o una alarma audible que indica el final de una fase de inyección. Baxter ha publicado una nueva versión de SIGMA Spectrum Infusion System, la 8, que incluye cambios en el software y el hardware.

*Credits: N/A
CVSS Scores
Attack Vector
Physical
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High
Attack Vector
Local
Attack Complexity
Low
Authentication
None
Confidentiality
Partial
Integrity
Partial
Availability
Partial
* Common Vulnerability Scoring System
SSVC
  • Decision:-
Exploitation
-
Automatable
-
Tech. Impact
-
* Organization's Worst-case Scenario
Timeline
  • 2014-08-22 CVE Reserved
  • 2019-03-26 CVE Published
  • 2023-03-08 EPSS Updated
  • 2024-08-06 CVE Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-259: Use of Hard-coded Password
  • CWE-798: Use of Hard-coded Credentials
CAPEC
References (1)
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Baxter
Search vendor "Baxter"
Sigma Spectrum Infusion System Firmware
Search vendor "Baxter" for product "Sigma Spectrum Infusion System Firmware"
6.05
Search vendor "Baxter" for product "Sigma Spectrum Infusion System Firmware" and version "6.05"
-
Affected
in Baxter
Search vendor "Baxter"
Sigma Spectrum Infusion System
Search vendor "Baxter" for product "Sigma Spectrum Infusion System"
--
Safe
Baxter
Search vendor "Baxter"
Sigma Spectrum Infusion System Firmware
Search vendor "Baxter" for product "Sigma Spectrum Infusion System Firmware"
6.05
Search vendor "Baxter" for product "Sigma Spectrum Infusion System Firmware" and version "6.05"
-
Affected
in Baxter
Search vendor "Baxter"
Wireless Battery Module
Search vendor "Baxter" for product "Wireless Battery Module"
16
Search vendor "Baxter" for product "Wireless Battery Module" and version "16"
-
Safe