// For flags

CVE-2024-48966

Life2000 service tools for test and calibration do not support user authentication

Severity Score

10.0
*CVSS v3.1

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

Attend
*SSVC
Descriptions

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

*Credits: N/A
CVSS Scores
Attack Vector
Network
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High
* Common Vulnerability Scoring System
SSVC
  • Decision:Attend
Exploitation
None
Automatable
Yes
Tech. Impact
Total
* Organization's Worst-case Scenario
Timeline
  • 2024-10-10 CVE Reserved
  • 2024-11-14 CVE Published
  • 2024-11-15 CVE Updated
  • 2024-11-15 EPSS Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-306: Missing Authentication for Critical Function
CAPEC
  • CAPEC-117: Interception
  • CAPEC-441: Malicious Logic Insertion
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Baxter
Search vendor "Baxter"
Life2000 Ventilation System
Search vendor "Baxter" for product "Life2000 Ventilation System"
06.08.00.00
Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00"
en
Affected