CVE-2024-48966
Life2000 service tools for test and calibration do not support user authentication
Severity Score
10.0
*CVSS v3.1
Exploit Likelihood
*EPSS
Affected Versions
*CPE
Public Exploits
0
*Multiple Sources
Exploited in Wild
-
*KEV
Decision
Attend
*SSVC
Descriptions
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
*Credits:
N/A
CVSS Scores
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
* Common Vulnerability Scoring System
SSVC
- Decision:Attend
Exploitation
Automatable
Tech. Impact
* Organization's Worst-case Scenario
Timeline
- 2024-10-10 CVE Reserved
- 2024-11-14 CVE Published
- 2024-11-15 CVE Updated
- 2024-11-15 EPSS Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-306: Missing Authentication for Critical Function
CAPEC
- CAPEC-117: Interception
- CAPEC-441: Malicious Logic Insertion
References (1)
URL | Date | SRC |
---|
URL | Date | SRC |
---|
URL | Date | SRC |
---|
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Baxter Search vendor "Baxter" | Life2000 Ventilation System Search vendor "Baxter" for product "Life2000 Ventilation System" | 06.08.00.00 Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00" | en |
Affected
|