Life2000 ventilator and Service PC lack sufficient audit logging capabilities

Severity Score

10.0
*CVSS v3.1

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

Attend
*SSVC
Descriptions

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

*Credits: N/A
CVSS Scores
Attack Vector
Network
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High
* Common Vulnerability Scoring System
SSVC
  • Decision:Attend
Exploitation
None
Automatable
Yes
Tech. Impact
Total
* Organization's Worst-case Scenario
Timeline
  • 2024-10-10 CVE Reserved
  • 2024-11-14 CVE Published
  • 2024-11-15 CVE Updated
  • 2024-11-15 EPSS Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-778: Insufficient Logging
CAPEC
  • CAPEC-117: Interception
  • CAPEC-441: Malicious Logic Insertion
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Baxter
Search vendor "Baxter"
Life2000 Ventilation System
Search vendor "Baxter" for product "Life2000 Ventilation System"
06.08.00.00
Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00"
en
Affected