CVE-2024-48967
Life2000 ventilator and Service PC lack sufficient audit logging capabilities
Severity Score
10.0
*CVSS v3.1
Exploit Likelihood
*EPSS
Affected Versions
*CPE
Public Exploits
0
*Multiple Sources
Exploited in Wild
-
*KEV
Decision
Attend
*SSVC
Descriptions
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
*Credits:
N/A
CVSS Scores
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
* Common Vulnerability Scoring System
SSVC
- Decision:Attend
Exploitation
Automatable
Tech. Impact
* Organization's Worst-case Scenario
Timeline
- 2024-10-10 CVE Reserved
- 2024-11-14 CVE Published
- 2024-11-15 CVE Updated
- 2024-11-15 EPSS Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-778: Insufficient Logging
CAPEC
- CAPEC-117: Interception
- CAPEC-441: Malicious Logic Insertion
References (1)
URL | Date | SRC |
---|
URL | Date | SRC |
---|
URL | Date | SRC |
---|
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Baxter Search vendor "Baxter" | Life2000 Ventilation System Search vendor "Baxter" for product "Life2000 Ventilation System" | 06.08.00.00 Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00" | en |
Affected
|