Debug port on Life2000 Ventilator serial interface is enabled by default

Severity Score

9.3
*CVSS v3.1

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

Track*
*SSVC
Descriptions

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

*Credits: N/A
CVSS Scores
Attack Vector
Local
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High
* Common Vulnerability Scoring System
SSVC
  • Decision:Track*
Exploitation
None
Automatable
No
Tech. Impact
Total
* Organization's Worst-case Scenario
Timeline
  • 2024-10-10 CVE Reserved
  • 2024-11-14 CVE Published
  • 2024-11-15 EPSS Updated
  • 2024-11-18 CVE Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-1263: Improper Physical Access Control
CAPEC
  • CAPEC-117: Interception
  • CAPEC-441: Malicious Logic Insertion
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Baxter
Search vendor "Baxter"
Life2000 Ventilation System
Search vendor "Baxter" for product "Life2000 Ventilation System"
06.08.00.00
Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00"
en
Affected