CVE-2024-48973
Debug port on Life2000 Ventilator serial interface is enabled by default
Severity Score
9.3
*CVSS v3.1
Exploit Likelihood
*EPSS
Affected Versions
*CPE
Public Exploits
0
*Multiple Sources
Exploited in Wild
-
*KEV
Decision
Track*
*SSVC
Descriptions
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
*Credits:
N/A
CVSS Scores
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
* Common Vulnerability Scoring System
SSVC
- Decision:Track*
Exploitation
Automatable
Tech. Impact
* Organization's Worst-case Scenario
Timeline
- 2024-10-10 CVE Reserved
- 2024-11-14 CVE Published
- 2024-11-15 EPSS Updated
- 2024-11-18 CVE Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-1263: Improper Physical Access Control
CAPEC
- CAPEC-117: Interception
- CAPEC-441: Malicious Logic Insertion
References (1)
URL | Date | SRC |
---|
URL | Date | SRC |
---|
URL | Date | SRC |
---|
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Baxter Search vendor "Baxter" | Life2000 Ventilation System Search vendor "Baxter" for product "Life2000 Ventilation System" | 06.08.00.00 Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00" | en |
Affected
|