// For flags

CVE-2024-48974

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Severity Score

9.3
*CVSS v3.1

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

Track*
*SSVC
Descriptions

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

*Credits: N/A
CVSS Scores
Attack Vector
Local
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High
* Common Vulnerability Scoring System
SSVC
  • Decision:Track*
Exploitation
None
Automatable
No
Tech. Impact
Total
* Organization's Worst-case Scenario
Timeline
  • 2024-10-10 CVE Reserved
  • 2024-11-14 CVE Published
  • 2024-11-15 EPSS Updated
  • 2024-11-18 CVE Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-494: Download of Code Without Integrity Check
CAPEC
  • CAPEC-117: Interception
  • CAPEC-186: Malicious Software Update
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Baxter
Search vendor "Baxter"
Life2000 Ventilation System
Search vendor "Baxter" for product "Life2000 Ventilation System"
06.08.00.00
Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00"
en
Affected