CVE-2024-48974
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Severity Score
9.3
*CVSS v3.1
Exploit Likelihood
*EPSS
Affected Versions
*CPE
Public Exploits
0
*Multiple Sources
Exploited in Wild
-
*KEV
Decision
Track*
*SSVC
Descriptions
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
*Credits:
N/A
CVSS Scores
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
* Common Vulnerability Scoring System
SSVC
- Decision:Track*
Exploitation
Automatable
Tech. Impact
* Organization's Worst-case Scenario
Timeline
- 2024-10-10 CVE Reserved
- 2024-11-14 CVE Published
- 2024-11-15 EPSS Updated
- 2024-11-18 CVE Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-494: Download of Code Without Integrity Check
CAPEC
- CAPEC-117: Interception
- CAPEC-186: Malicious Software Update
References (1)
URL | Date | SRC |
---|
URL | Date | SRC |
---|
URL | Date | SRC |
---|
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Baxter Search vendor "Baxter" | Life2000 Ventilation System Search vendor "Baxter" for product "Life2000 Ventilation System" | 06.08.00.00 Search vendor "Baxter" for product "Life2000 Ventilation System" and version "06.08.00.00" | en |
Affected
|