CVE-2025-0626
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Severity Score
Exploit Likelihood
Affected Versions
Public Exploits
0Exploited in Wild
-Decision
Descriptions
The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
This fact sheet details an analysis of three firmware package versions of the Contec CMS8000, a patient monitor used by the U.S. Healthcare and Public Health (HPH) sector. Analysts discovered that an embedded backdoor function with a hard-coded IP address and functionality that enables patient data spillage exists in all versions analyzed.
CVSS Scores
SSVC
- Decision:Track*
Timeline
- 2025-01-21 CVE Reserved
- 2025-01-30 CVE Published
- 2025-03-01 CVE Updated
- 2025-04-15 EPSS Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-912: Hidden Functionality
CAPEC
References (2)
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Contec Health Search vendor "Contec Health" | CMS8000 Patient Monitor Search vendor "Contec Health" for product "CMS8000 Patient Monitor" | * | en |
Affected
|