CVE-2025-0683
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Severity Score
Exploit Likelihood
Affected Versions
Public Exploits
0Exploited in Wild
-Decision
Descriptions
In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
This fact sheet details an analysis of three firmware package versions of the Contec CMS8000, a patient monitor used by the U.S. Healthcare and Public Health (HPH) sector. Analysts discovered that an embedded backdoor function with a hard-coded IP address and functionality that enables patient data spillage exists in all versions analyzed.
CVSS Scores
SSVC
- Decision:-
Timeline
- 2025-01-23 CVE Reserved
- 2025-01-30 CVE Published
- 2025-01-31 CVE Updated
- 2025-01-31 EPSS Updated
- ---------- Exploited in Wild
- ---------- KEV Due Date
- ---------- First Exploit
CWE
- CWE-359: Exposure of Private Personal Information to an Unauthorized Actor
CAPEC
References (2)
Affected Vendors, Products, and Versions
Vendor | Product | Version | Other | Status | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Vendor | Product | Version | Other | Status | <-- --> | Vendor | Product | Version | Other | Status |
Contec Health Search vendor "Contec Health" | CMS8000 Patient Monitor Search vendor "Contec Health" for product "CMS8000 Patient Monitor" | * | en |
Affected
|