// For flags

CVE-2025-0683

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

Severity Score

8.2
*CVSS v4

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

-
*SSVC
Descriptions

In its default configuration, the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

This fact sheet details an analysis of three firmware package versions of the Contec CMS8000, a patient monitor used by the U.S. Healthcare and Public Health (HPH) sector. Analysts discovered that an embedded backdoor function with a hard-coded IP address and functionality that enables patient data spillage exists in all versions analyzed.

*Credits: An anonymous researcher reported these vulnerabilities to CISA.
CVSS Scores
Attack Vector
Network
Attack Complexity
Low
Attack Requirements
Present
Privileges Required
None
User Interaction
None
System
Vulnerable | Subsequent
Confidentiality
High
None
Integrity
None
None
Availability
None
None
Attack Vector
Network
Attack Complexity
High
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
None
Availability
None
Attack Vector
Network
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
None
Availability
None
Attack Vector
Network
Attack Complexity
Low
Authentication
None
Confidentiality
Partial
Integrity
None
Availability
None
* Common Vulnerability Scoring System
SSVC
  • Decision:-
Exploitation
-
Automatable
-
Tech. Impact
-
* Organization's Worst-case Scenario
Timeline
  • 2025-01-23 CVE Reserved
  • 2025-01-30 CVE Published
  • 2025-01-31 CVE Updated
  • 2025-01-31 EPSS Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-359: Exposure of Private Personal Information to an Unauthorized Actor
CAPEC
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Contec Health
Search vendor "Contec Health"
CMS8000 Patient Monitor
Search vendor "Contec Health" for product "CMS8000 Patient Monitor"
*en
Affected