// For flags

CVE-2025-1204

 

Severity Score

7.7
*CVSS v4

Exploit Likelihood

*EPSS

Affected Versions

*CPE

Public Exploits

0
*Multiple Sources

Exploited in Wild

-
*KEV

Decision

Attend
*SSVC
Descriptions

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

*Credits: Claroty - Team82
CVSS Scores
Attack Vector
Network
Attack Complexity
Low
Attack Requirements
Present
Privileges Required
None
User Interaction
Passive
System
Vulnerable | Subsequent
Confidentiality
High
None
Integrity
High
None
Availability
High
None
Attack Vector
Network
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
High
Availability
High
Attack Vector
Network
Attack Complexity
Low
Authentication
None
Confidentiality
None
Integrity
Partial
Availability
Partial
* Common Vulnerability Scoring System
SSVC
  • Decision:Attend
Exploitation
Poc
Automatable
No
Tech. Impact
Total
* Organization's Worst-case Scenario
Timeline
  • 2025-02-10 CVE Reserved
  • 2025-02-25 CVE Published
  • 2025-02-25 CVE Updated
  • 2025-05-04 EPSS Updated
  • ---------- Exploited in Wild
  • ---------- KEV Due Date
  • ---------- First Exploit
CWE
  • CWE-912: Hidden Functionality
CAPEC
Affected Vendors, Products, and Versions
Vendor Product Version Other Status
Vendor Product Version Other Status <-- --> Vendor Product Version Other Status
Contec Health
Search vendor "Contec Health"
CMS8000 Patient Monitor
Search vendor "Contec Health" for product "CMS8000 Patient Monitor"
0
Search vendor "Contec Health" for product "CMS8000 Patient Monitor" and version "0"
en
Affected