8 results (0.005 seconds)

CVSS: 6.1EPSS: 0%CPEs: 3EXPL: 0

The use of a hard-coded cryptographic key significantly increases the possibility encrypted data may be recovered from the Patient Information Center iX (PIC iX) Versions B.02, C.02, and C.03. El uso de una clave criptográfica embebida aumenta significativamente la posibilidad de que los datos encriptados puedan ser recuperados de Patient Information Center iX (PIC iX) Versiones B.02, C.02 y C.03 • https://www.cisa.gov/uscert/ics/advisories/icsma-21-322-02 • CWE-321: Use of Hard-coded Cryptographic Key •

CVSS: 6.8EPSS: 0%CPEs: 3EXPL: 0

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges. Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versión A.01, Monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores. El producto expone un recurso a la esfera de control equivocada, proporcionando a los actores no deseados un acceso inapropiado al recurso. La aplicación en la estación de supervisión opera en modo Kiosk, que es vulnerable a filtraciones locales que podrían permitir a un atacante con acceso físico escapar el entorno restringido con privilegios limitados • https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 https://www.philips.com/productsecurity • CWE-668: Exposure of Resource to Wrong Sphere •

CVSS: 4.3EPSS: 0%CPEs: 4EXPL: 0

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling. Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versión A.01, Monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores. El producto recibe una entrada que se espera que esté bien formada (es decir, que cumpla con una determinada sintaxis) pero no comprueba o comprueba incorrectamente que la entrada cumple con la sintaxis, causando que el servicio de inscripción de certificados se bloque. No impacta la supervisión, pero evita que se inscriban nuevos dispositivos • https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 https://www.philips.com/productsecurity • CWE-1286: Improper Validation of Syntactic Correctness of Input •

CVSS: 6.5EPSS: 0%CPEs: 26EXPL: 0

In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart. Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versión A.01, Monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores. El producto recibe una entrada o datos pero no comprueba o comprueba incorrectamente que la entrada presenta las propiedades necesarias para procesar los datos seguramente y correctamente, lo que puede inducir una condición de denegación de servicio por medio de un reinicio del sistema • https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 https://www.philips.com/productsecurity • CWE-20: Improper Input Validation •

CVSS: 6.4EPSS: 0%CPEs: 26EXPL: 0

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate. Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versión A.01, Monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores. El software no comprueba o comprueba inapropiadamente el estado de revocación de un certificado, lo que puede causar que use un certificado comprometido • https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 https://www.philips.com/productsecurity • CWE-299: Improper Check for Certificate Revocation •